HOME  DEUTSCH  FRANÇAIS
red-y for gasflow Voegtlin Instruments AG
red-y smart series
           
  Technical overview 'red-y smart series'  
           
smart series:
» Overview
» Instrument Types
» Technical Data
» PDF-Datasheet
» Contact
Instrument versions   standard hi-performance  
  Accuracy ± 1,5% of full scale ± 0,3% of full scale and ± 0,5% of reading
  Turndown ratio 1 : 30 1 : 100 (higher on request)
     
Performance data Media Air, O2, N2, He, Ar, CO2, H2, CH4, C3H8, SF6
    Other gases and gas mixtures on request
(real gas calibration or conversion factors)
  Response time 50ms
  Repeatability ± 0.2% of full scale
  Longterm stability < 1% of measured value / year
  Power supply 24 Vdc (18 – 30 Vdc)
  Current drawn Meter: max. 100mA
    Controller: max. 250mA
  Operation pressure 0.4 - 11 bar a
  Temperature 0 – 50°C
  Materials Anodized aluminium, optional stainless steel electropolished
  Seals FKM, NBR (valve), optional EPDM
  Pressure sensitivity < 0.2%/bar
  Temperature sensitivity < 0.025% / °C
     
Measuring Ranges Type Measuring range (Air), Full scale freely selectable Connection
red-y smart meter GSM-A from 0 ... 25 mln/min to 0 ... 500 mln/min G1/4"
  GSM-B from 0 ... 500 mln/min to 0 ... 5000 mln/min G1/4"
  GSM-C from 0 ... 5 ln/min to 0 ... 50 ln/min G1/4"
  GSM-D from 0 ... 50 ln/min to 0 ... 450 ln/min G1/2"
red-y smart controller GSC-A from 0 ... 25 mln/min to 0 ... 500 mln/min G1/4"
  GSC-B from 0 ... 500 mln/min to 0 ... 5000 mln/min G1/4"
  GSC-C from 0 ... 5 ln/min to 0 ... 50 ln/min G1/4"
  GSC-D from 0 ... 50 ln/min to 0 ... 450 ln/min G1/2"
     
Integration Output signals:  
  - analog 0..20 mA, 4..20 mA, 0..5 V, 1..5 V, 0..10 V, 2..10 V
  - digital RS-485; Modbus RTU (Slave); Lab View-VI‘s available
Optional: ProfiBus DP-V0, DP-V1
  Process connection G¼“ female less than 50 ln/min, G½“ female less than 450 ln/min
  Inlet section None required
  Electrical connection Sub D plug, 9 pole
  Mounting orientation Any orientation (horizontal only above 5 bar)
     
Safety Test pressure 16 bar a
  Leak rate < 1 x 10-6 mbar l/s He
  Environ. protection IP-50
  EMC EN 61326-1
     
Software 'get red-y' View flow rate & temperature – Graphical display of measured data
  Change setpoints – Select measured gas and mixtures
  Option: Datalogging, Gasmixer
  > More Information about PC-Software 'get red-y'